IMAGE

Fig. 3.

ID
ZDB-IMAGE-230211-81
Source
Figures for Hajiabadi et al., 2022
Image
Figure Caption

Fig. 3.

A two-phase acquisition protocol to combine acquisition of quality control images with high-speed time-lapse imaging. (A) Image data were acquired at multiple positions in a sample, thus obtaining multiple viewpoints containing several objects of interest (nuclei, indicated as circles). (B) For each position, a sequence of two acquisition phases is carried out. In phase A, for each z position, a low-quality image, two high-quality reference images, and two low-quality test images are recorded. Low-quality images are recorded at a shortened exposure time (texp), high-quality images at a reference exposure time resulting in images of the desired quality (tref). Acquisition phase A obtains the images required for n2v model training as well as the assessment of effective image resolution and reconstruction errors. In phase B, only single low-quality images are recorded with the shortened exposure time (texp), resulting in an increased rate of acquisition compared to acquisition with full exposure time (tref). Acquisition phase B obtains only low-quality images, which are reconstructed after the experiment is completed.

Acknowledgments
This image is the copyrighted work of the attributed author or publisher, and ZFIN has permission only to display this image to its users. Additional permissions should be obtained from the applicable author or publisher of the image. Full text @ PNAS Nexus